Technology/Device Do-It-Yourself Open Artificial Pancreas System in Children and Adolescents with Type 1 Diabetes Mellitus: Real-World Data
Min Sun Choi1*orcid, Seunghyun Lee2*orcid, Jiwon Kim3, Gyuri Kim1, Sung Min Park2,3orcidcorresp_icon, Jae Hyeon Kim1,4orcidcorresp_icon
Diabetes & Metabolism Journal 2022;46(1):154-159.
DOI: https://doi.org/10.4093/dmj.2021.0011
Published online: November 23, 2021
  • 6,333 Views
  • 214 Download
  • 5 Web of Science
  • 6 Crossref
  • 7 Scopus

1Division of Endocrinology and Metabolism, Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea

2Department of Creative IT Engineering, Pohang University of Science and Technology (POSTECH), Pohang, Korea

3Department of Electrical Engineering, Pohang University of Science and Technology (POSTECH), Pohang, Korea

4Department of Clinical Research Design & Evaluation, Samsung Advanced Institute for Health Sciences & Technology, Sungkyunkwan University, Seoul, Korea

corresp_icon Corresponding authors: Sung Min Park orcid Department of Creative IT Engineering, Pohang University of Science and Technology (POSTECH), 77 Cheongam-ro, Nam-gu, Pohang 37673, Korea E-mail: sungminpark@postech.ac.kr
Jae Hyeon Kim orcid Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, 81 Irwon-ro, Gangnam-gu, Seoul 06351, Korea E-mail: jaehyeon@skku.edu
*Min Sun Choi and Seunghyun Lee contributed equally to this study as first authors.
• Received: January 20, 2021   • Accepted: May 20, 2021

Copyright © 2022 Korean Diabetes Association

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

prev next
  • Few studies have been conducted among Asian children and adolescents with type 1 diabetes mellitus (T1DM) using do-it-yourself artificial pancreas system (DIY-APS). We evaluated real-world data of pediatric T1DM patients using DIY-APS. Data were obtained for 10 patients using a DIY-APS with algorithms. We collected sensor glucose and insulin delivery data from each participant for a period of 4 weeks. Average glycosylated hemoglobin was 6.2%±0.3%. The mean percentage of time that glucose level remained in the target range of 70 to 180 mg/dL was 82.4%±7.8%. Other parameters including time above range, time below range and mean glucose were also within the recommended level, similar to previous commercial and DIY-APS studies. However, despite meeting the target range, unadjusted gaps were still observed between the median basal setting and temporary basal insulin, which should be handled by healthcare providers.
With the demand for strict glycemic control and advances in diabetes technology, the use of automated insulin delivery systems known as artificial pancreas system (APS) has emerged. Previous studies showed that APS improves glycosylated hemoglobin (HbA1c) level and the percentage of time blood glucose level is within 70 to 180 mg/dL (time in range [TIR]) while preventing hypoglycemia in patients with type 1 diabetes mellitus (T1DM) [1-3]. However, most APS have not been approved by the U.S. Food and Drug Administration, except for the MiniMed 670G (Medtronic Inc., Dublin, Ireland) and Control-IQ (Tandem Diabetes Care, San Diego, CA, USA) systems, which are not available in Korea. Therefore, do-it-yourself APS (DIY-APS) has been used by patients who desire to construct loop systems using open-source algorithms [4]. Few DIY-APS studies have been conducted in Asia. In addition, few studies have provided detailed graphs showing glucose and insulin profiles for individuals, with the exception of a Czech pilot study [5]. In this observational study, we investigated real-world DIY-APS data of children and adolescents with T1DM in Korea.
Participant recruitment
We posted a recruitment notice on the website of the Korean Society of Type 1 Diabetes (http://www.ksT1DM.org/). We recruited subjects with T1DM who had been using DIY-APS for more than 1 month. The subjects voluntarily participated in the study and provided written informed consent. The study protocol was approved by the Institutional Review Board (IRB) of Samsung Medical Center (IRB-No 2018-08-142) and was performed according to the guidelines of the Declaration of Helsinki.
Data acquisition
Patient blood glucose and insulin data were collected through a continuous glucose monitoring (CGM) and an insulin infusion device. Data were stored in the cloud or to a mobile phone, the same as before the study participation. The cloud was modified to transmit uploaded basal insulin data to our research server in real time. The cloud was rolled back to its pre-experimental version at the end of data acquisition. Demographic information including age, sex, height, body weight, the level of HbA1c, duration of diabetes and APS use were obtained.
Data analysis
The CGM parameters in both the daytime (6:00 AM to 12:00 AM) and nighttime (12:00 AM to 6:00 AM) were assessed, including TIR, the percentage of time blood glucose level was above 180 mg/dL (time above range [TAR]) and below 70 mg/dL (time below range [TBR]), percent coefficient of variation (%CV) and glucose management indicator (GMI). The number of hypoglycemic events, basal and bolus insulin doses, basal changes, and bolus injections were assessed. We described average blood glucose values in the form of ambulatory glucose profile reports and insulin delivery graphs of each individual. Statistical analyses were performed using SPSS version 25.0 (IBM Corp., Armonk, NY, USA).
Among 28 screened subjects who had been using DIY-APS, a total of 11 patients were included in the study (Supplementary Fig. 1). However, since case no. 11 used an open loop as sensor-augmented pump rather than a closed loop with a DIY-APS algorithm (Fig. 1), we omitted case no. 11 from analysis.
The clinical characteristics and results of 10 DIY-APS users are described (Table 1). All patients were less than 19 years of age, with an average age of 10.1±3.2 years. The average HbA1c level was 6.2%±0.3%. The median duration of diabetes was 3.1 years and the median duration of DIY-APS usage was 1.9 years. Among all patients, mean TIR was 82.4%±7.8% over the total time. Nighttime TIR (84.14%±6.78%) was higher than daytime TIR (82.31%±9.0%). TBR of <70 mg/dL (2.42%) and <54 mg/dL (0.65%) were less than the recommended clinical targets [6]. During daytime, the average glucose level, frequency of basal rate change, bolus insulin injection and hypoglycemic events were significantly higher than in the nighttime.
Graphs of insulin delivery and blood glucose levels of each individual are shown (Fig. 1, Supplementary Fig. 2). The red line in the graph represents the basal setting for each patient. Case no. 7 achieved the lowest CV (26.4%) and highest TIR (91.01) among the participants. Case no. 8 showed skewed patterns of median temporary basal values and high CV. Unadjusted gaps were seen between the median basal setting and temporary basal setting, despite the appropriate mean glucose or GMI.
Results for TIR, TAR, TBR and mean glucose satisfied the recommended level [6] set in previous studies using commercial APS such as the MiniMed 670G or Control-IQ system [2,7].
To the best of our knowledge, this is the first study reporting CGM data along with a graph of delivered insulin dose for each individual with T1DM. We observed that overall CGM variables were within the recommended target ranges throughout the daytime and nighttime [6,8]. Furthermore, we found that some patients used DIY-APS without correcting the basal insulin preset, even though the preset was outside the 10 to 90 percentile of the temporary basal range. Unadjusted settings existed even when TIR, GMI and CV were appropriate. This finding reinforces the role of healthcare providers in helping patients adjust their DIY-APS settings [9].
Higher TIR and less frequent hypoglycemic events were observed in the nighttime than in daytime, while the temporary basal setting was often a negative value. These findings suggest that the basal insulin settings were too strict, especially during the daytime, probably with the intention of strict glucose control.
We compared the data with those of previous studies of pediatric patients using APS. Braune et al. [10] reported that DIY-APS improved HbA1c levels and TIR in patients with well-controlled glycemic status before use of DIY-APS. Breton et al. [7] showed that pediatric T1DM patients using the Tandem Control-IQ system achieved better glycemic control than sensor-augmented insulin pump users. In line with previous studies, our study reconfirmed that appropriate glycemic control could be achieved in pediatric DIY-APS users without increasing hypoglycemia. However, since the participants were highly motivated, promising TIR after using DIY-APS could be associated with a well-controlled TIR before using DIY-APS [11]. Therefore, head-to-head studies comparing APS algorithms are needed. A recent multinational randomized controlled trial reported an investigational advanced hybrid closed-loop system reduced hyperglycemia without increasing hypoglycemia in adolescents and young adults with T1DM compared with MiniMed 670G [12].
Many patients have experienced improved quality of life after using DIY-APS [4,13]. Reduction of HbA1c, increased TIR and less hypoglycemia gave them confidence in managing their diabetes and led to increased feelings of safety while sleeping or engaging in physical activities.
This study has several limitations. First, selection bias existed due to the study design. It was impossible to confirm whether the use of DIY-APS improved glycemic control compared to before DIY-APS initiation. Socioeconomic status might also have effects on accessibility to DIY-APS. Second, due to the small sample size, it is difficult to apply the outcomes to other children/adult patients with T1DM using DIY-APS. Large-scale randomized studies are needed, including more children/adult patients with T1DM using DIY-APS [14]. Last, the latest APS features such as Autosens, advanced meal assist, super microbolus (SMB), and meal announcements were not available to most of participants. Some caregivers avoided using SMB due to safety concerns and performed manual bolus injections instead of using a bolus calculator. For frequent manual bolus users, the recently released Control-IQ algorithm, which has an automated correction bolus, might be helpful to reduce manual bolus injections.
In conclusion, it was possible to achieve appropriate glycemic control in pediatric T1DM patients in real-world use of DIY-APS. Healthcare providers should monitor the DIY-APS settings of each patient even when CGM parameters are within target glucose ranges.
Supplementary materials related to this article can be found online at https://doi.org/10.4093/dmj.2021.0011.
Supplementary Fig. 1.
Patient disposition. APS, artificial pancreas system.
dmj-2021-0011-suppl1.pdf
Supplementary Fig. 2.
Graphs of insulin delivery and blood glucose levels for residual cases. GMI, glucose management indicator; CV, coefficient of variation; TDD, total daily dose of insulin.
dmj-2021-0011-suppl2.pdf

CONFLICTS OF INTEREST

No potential conflict of interest relevant to this article was reported.

AUTHOR CONTRIBUTIONS

Conception or design: M.S.C., S.L., S.M.P., J.H.K.

Acquisition, analysis, or interpretation of data: M.S.C., S.L., J. K., G.K., S.M.P., J.H.K.

Drafting the work or revising: M.S.C., S.L., S.M.P., J.H.K.

Final approval of the manuscript: M.S.C., S.L., J.K., G.K., S.M.P., J.H.K.

FUNDING

None

Acknowledgements
We would like to thank the DIY-APS users in the Korean Society of Type 1 Diabetes (SugarTree, http://www.ksT1DM.org/) who shared their DIY-APS records with us for the current study. This research did not receive any support or funding from the public or commercial sectors.
Fig. 1.
Graphs of insulin delivery and blood glucose levels for representative cases. GMI, glucose management indicator; CV, coefficient of variation; TDD, total daily dose of insulin.
dmj-2021-0011f1.jpg
Table 1.
Clinical characteristics and summary results of 4 weeks of DIY-APS usage
Variable Overall Daytime (6:00 AM to 12:00 AM) Nighttime (12:00 AM to 6:00 AM) P value
No. of participants 10 - - -
Age, yr 10.1±3.2 - - -
Male/Female 7 (70.0)/3 (30.0) - - -
Weight, kg 33.0±11.9 - - -
BMI, kg/m2 17.2±2.4 - - -
Duration of diabetes, yr 3.1 (1.8 to 6.5) - - -
HbA1c, % 6.2±0.3 - - -
Duration of DIY-APS usage, yr 1.9±1.1 - - -
Average glucose level, mg/dL 135.8±11.92 137.78±12.48 130.01±13.74 0.017
% CV of glucose 31.57±4.43 31.43±4.7 28.46 (27.05 to 33.75) 0.799
TAR >250 mg/dL, % 1.01 (0.65 to 2.39) 1.18 (0.73 to 2.61) 0.79 (0.08 to 2.49) 0.333
TAR >180 mg/dL, % 14.77±7.85 14.34 (11.96 to 16.55) 12.16±7.08 0.093
TIR 70–180 mg/dL, % 82.8±8.12 82.31±9.06 84.14± 6.78 0.799
TBR <70 mg/dL, % 2.42±1.39 2.0±1.27 3.7±3.33 0.074
TBR <54 mg/dL, % 0.65±0.5 0.45 (0.13 to 1.01) 0.49 (0.19 to 1.66) 0.445
Hypoglycemic events
 <70 mg/dL, n/day 0.76±0.41 0.53±0.32 0.24±0.15 0.011
 <54 mg/dL, n/day 0.21±0.16 0.11 (0.08 to 0.28) 0.04 (0.01 to 0.09) 0.035
 Prolonged, n/day 0 (0 to 0.03) 0 0 1.000
GMI, % 6.56±0.29 6.61±0.3 6.42±0.33 0.017
TDD, unit/kg/day 0.79±0.16 0.66 (0.62 to 0.73) 0.09±0.02 0.005
 Basal, % 42.25±7.57 37.07±7.88 86.16±8.91 0.005
 Bolus, % 57.75±7.57 62.93±7.88 13.84±8.91 0.005
Median temporary basal
 Per hour, % –21.42 (–23.25 to –17.92) –23.44±8.98 –10.65±25.32 0.386
 Per day, % –10.94±11.1
Frequency of basal changes, n/hr 2.05 (1.97 to 2.13) 2.26 (2.19 to 2.37) 1.58 (1.52 to 1.72) 0.005
Bolus, n/day 11.8±3.12 11.1±3.08 0.5 (0.43 to 0.88) 0.005

Values are presented as mean±standard deviation, number (%), or median (interquartile range).

DIY, do-it-yourself; APS, artificial pancreas system; BMI, body mass index; HbA1c, glycosylated hemoglobin; CV, coefficient of variation; TAR, time above range; TIR, time in range; TBR, time below range; GMI, glucose management indicator; TDD, total daily insulin dose.

  • 1. Sherr JL, Cengiz E, Palerm CC, Clark B, Kurtz N, Roy A, et al. Reduced hypoglycemia and increased time in target using closed-loop insulin delivery during nights with or without antecedent afternoon exercise in type 1 diabetes. Diabetes Care 2013;36:2909-14.ArticlePubMedPMCPDF
  • 2. Garg SK, Weinzimer SA, Tamborlane WV, Buckingham BA, Bode BW, Bailey TS, et al. Glucose outcomes with the in-home use of a hybrid closed-loop insulin delivery system in adolescents and adults with type 1 diabetes. Diabetes Technol Ther 2017;19:155-63.ArticlePubMedPMC
  • 3. Melmer A, Zuger T, Lewis DM, Leibrand S, Stettler C, Laimer M. Glycaemic control in individuals with type 1 diabetes using an open source artificial pancreas system (OpenAPS). Diabetes Obes Metab 2019;21:2333-7.ArticlePubMedPDF
  • 4. Lewis D, LeibrandS S; #OpenAPS Community. Real-world use of open source artificial pancreas systems. J Diabetes Sci Technol 2016;10:1411.ArticlePubMedPMCPDF
  • 5. Petruzelkova L, Soupal J, Plasova V, Jiranova P, Neuman V, Plachy L, et al. Excellent glycemic control maintained by open-source hybrid closed-loop androidAPS during and after sustained physical activity. Diabetes Technol Ther 2018;20:744-50.ArticlePubMed
  • 6. American Diabetes Association. 6. Glycemic targets: standards of medical care in diabetes-2021. Diabetes Care 2021;44(Suppl 1):S73-84.ArticlePubMedPDF
  • 7. Breton MD, Kanapka LG, Beck RW, Ekhlaspour L, Forlenza GP, Cengiz E, et al. A randomized trial of closed-loop control in children with type 1 diabetes. N Engl J Med 2020;383:836-45.ArticlePubMedPMC
  • 8. Aleppo G, Webb K. Continuous glucose monitoring integration in clinical practice: a stepped guide to data review and interpretation. J Diabetes Sci Technol 2019;13:664-73.ArticlePubMedPDF
  • 9. American Diabetes Association. 7. Diabetes technology: standards of medical care in diabetes-2020. Diabetes Care 2020;43(Suppl 1):S77-88.ArticlePubMedPDF
  • 10. Braune K, O’Donnell S, Cleal B, Lewis D, Tappe A, Willaing I, et al. Real-world use of do-it-yourself artificial pancreas systems in children and adolescents with type 1 diabetes: online survey and analysis of self-reported clinical outcomes. JMIR Mhealth Uhealth 2019;7:e14087.ArticlePubMedPMC
  • 11. Schoelwer MJ, Kanapka LG, Wadwa RP, Breton MD, Ruedy KJ, Ekhlaspour L, et al. Predictors of time-in-range (70-180mg/dL) achieved using a closed-loop control system. Diabetes Technol Ther 2021;23:475-81.ArticlePubMedPMC
  • 12. Bergenstal RM, Nimri R, Beck RW, Criego A, Laffel L, Schatz D, et al. A comparison of two hybrid closed-loop systems in adolescents and young adults with type 1 diabetes (FLAIR): a multicentre, randomised, crossover trial. Lancet 2021;397:208-19.ArticlePubMedPMC
  • 13. Ahmed SH, Ewins DL, Bridges J, Timmis A, Payne N, Mooney C, et al. Do-it-yourself (DIY) artificial pancreas systems for type 1 diabetes: perspectives of two adult users, parent of a user and healthcare professionals. Adv Ther 2020;37:3929-41.ArticlePubMedPMCPDF
  • 14. Burnside M, Lewis D, Crocket H, Wilson R, Williman J, Jefferies C, et al. CREATE (Community deRivEd AutomaTEd insulin delivery) trial: randomised parallel arm open label clinical trial comparing automated insulin delivery using a mobile controller (AnyDANA-loop) with an open-source algorithm with sensor augmented pump therapy in type 1 diabetes. J Diabetes Metab Disord 2020;19:1-15.ArticlePubMedPMCPDF

Figure & Data

References

    Citations

    Citations to this article as recorded by  
    • Real-world efficacy and safety of open-source automated insulin delivery for people with type 1 diabetes mellitus: Experience from mainland China
      Yongwen Zhou, Mengyun Lei, Daizhi Yang, Ping Ling, Ying Ni, Hongrong Deng, Wen Xu, Xubin Yang, Jinhua Yan, Benjamin John Wheeler, Jianping Weng
      Diabetes Research and Clinical Practice.2024; : 111910.     CrossRef
    • Advances in Continuous Glucose Monitoring and Integrated Devices for Management of Diabetes with Insulin-Based Therapy: Improvement in Glycemic Control
      Jee Hee Yoo, Jae Hyeon Kim
      Diabetes & Metabolism Journal.2023; 47(1): 27.     CrossRef
    • Open-source automated insulin delivery systems (OS-AIDs) in a pediatric population with type 1 diabetes in a real-life setting: the AWeSoMe study group experience
      Judith Nir, Marianna Rachmiel, Abigail Fraser, Yael Lebenthal, Avivit Brener, Orit Pinhas-Hamiel, Alon Haim, Eve Stern, Noa Levek, Tal Ben-Ari, Zohar Landau
      Endocrine.2023; 81(2): 262.     CrossRef
    • Efficacy and safety of Android artificial pancreas system use at home among adults with type 1 diabetes mellitus in China: protocol of a 26-week, free-living, randomised, open-label, two-arm, two-phase, crossover trial
      Mengyun Lei, Beisi Lin, Ping Ling, Zhigu Liu, Daizhi Yang, Hongrong Deng, Xubin Yang, Jing Lv, Wen Xu, Jinhua Yan
      BMJ Open.2023; 13(8): e073263.     CrossRef
    • Barriers to Uptake of Open-Source Automated Insulin Delivery Systems: Analysis of Socioeconomic Factors and Perceived Challenges of Caregivers of Children and Adolescents With Type 1 Diabetes From the OPEN Survey
      Antonia Huhndt, Yanbing Chen, Shane O’Donnell, Drew Cooper, Hanne Ballhausen, Katarzyna A. Gajewska, Timothée Froment, Mandy Wäldchen, Dana M. Lewis, Klemens Raile, Timothy C. Skinner, Katarina Braune
      Frontiers in Clinical Diabetes and Healthcare.2022;[Epub]     CrossRef
    • Toward Personalized Hemoglobin A1c Estimation for Type 2 Diabetes
      Namho Kim, Da Young Lee, Wonju Seo, Nan Hee Kim, Sung-Min Park
      IEEE Sensors Journal.2022; 22(23): 23023.     CrossRef

    • PubReader PubReader
    • ePub LinkePub Link
    • Cite this Article
      Cite this Article
      export Copy Download
      Close
      Download Citation
      Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

      Format:
      • RIS — For EndNote, ProCite, RefWorks, and most other reference management software
      • BibTeX — For JabRef, BibDesk, and other BibTeX-specific software
      Include:
      • Citation for the content below
      Do-It-Yourself Open Artificial Pancreas System in Children and Adolescents with Type 1 Diabetes Mellitus: Real-World Data
      Diabetes Metab J. 2022;46(1):154-159.   Published online November 23, 2021
      Close
    • XML DownloadXML Download
    Figure
    • 0
    Related articles
    Do-It-Yourself Open Artificial Pancreas System in Children and Adolescents with Type 1 Diabetes Mellitus: Real-World Data
    Image
    Fig. 1. Graphs of insulin delivery and blood glucose levels for representative cases. GMI, glucose management indicator; CV, coefficient of variation; TDD, total daily dose of insulin.
    Do-It-Yourself Open Artificial Pancreas System in Children and Adolescents with Type 1 Diabetes Mellitus: Real-World Data
    Variable Overall Daytime (6:00 AM to 12:00 AM) Nighttime (12:00 AM to 6:00 AM) P value
    No. of participants 10 - - -
    Age, yr 10.1±3.2 - - -
    Male/Female 7 (70.0)/3 (30.0) - - -
    Weight, kg 33.0±11.9 - - -
    BMI, kg/m2 17.2±2.4 - - -
    Duration of diabetes, yr 3.1 (1.8 to 6.5) - - -
    HbA1c, % 6.2±0.3 - - -
    Duration of DIY-APS usage, yr 1.9±1.1 - - -
    Average glucose level, mg/dL 135.8±11.92 137.78±12.48 130.01±13.74 0.017
    % CV of glucose 31.57±4.43 31.43±4.7 28.46 (27.05 to 33.75) 0.799
    TAR >250 mg/dL, % 1.01 (0.65 to 2.39) 1.18 (0.73 to 2.61) 0.79 (0.08 to 2.49) 0.333
    TAR >180 mg/dL, % 14.77±7.85 14.34 (11.96 to 16.55) 12.16±7.08 0.093
    TIR 70–180 mg/dL, % 82.8±8.12 82.31±9.06 84.14± 6.78 0.799
    TBR <70 mg/dL, % 2.42±1.39 2.0±1.27 3.7±3.33 0.074
    TBR <54 mg/dL, % 0.65±0.5 0.45 (0.13 to 1.01) 0.49 (0.19 to 1.66) 0.445
    Hypoglycemic events
     <70 mg/dL, n/day 0.76±0.41 0.53±0.32 0.24±0.15 0.011
     <54 mg/dL, n/day 0.21±0.16 0.11 (0.08 to 0.28) 0.04 (0.01 to 0.09) 0.035
     Prolonged, n/day 0 (0 to 0.03) 0 0 1.000
    GMI, % 6.56±0.29 6.61±0.3 6.42±0.33 0.017
    TDD, unit/kg/day 0.79±0.16 0.66 (0.62 to 0.73) 0.09±0.02 0.005
     Basal, % 42.25±7.57 37.07±7.88 86.16±8.91 0.005
     Bolus, % 57.75±7.57 62.93±7.88 13.84±8.91 0.005
    Median temporary basal
     Per hour, % –21.42 (–23.25 to –17.92) –23.44±8.98 –10.65±25.32 0.386
     Per day, % –10.94±11.1
    Frequency of basal changes, n/hr 2.05 (1.97 to 2.13) 2.26 (2.19 to 2.37) 1.58 (1.52 to 1.72) 0.005
    Bolus, n/day 11.8±3.12 11.1±3.08 0.5 (0.43 to 0.88) 0.005
    Table 1. Clinical characteristics and summary results of 4 weeks of DIY-APS usage

    Values are presented as mean±standard deviation, number (%), or median (interquartile range).

    DIY, do-it-yourself; APS, artificial pancreas system; BMI, body mass index; HbA1c, glycosylated hemoglobin; CV, coefficient of variation; TAR, time above range; TIR, time in range; TBR, time below range; GMI, glucose management indicator; TDD, total daily insulin dose.

    Choi MS, Lee S, Kim J, Kim G, Park SM, Kim JH. Do-It-Yourself Open Artificial Pancreas System in Children and Adolescents with Type 1 Diabetes Mellitus: Real-World Data. Diabetes Metab J. 2022;46(1):154-159.
    Received: Jan 20, 2021; Accepted: May 20, 2021
    DOI: https://doi.org/10.4093/dmj.2021.0011.

    Diabetes Metab J : Diabetes & Metabolism Journal
    Close layer
    TOP